Additional Function
Total measurement
time
Function (R, X, V, T)/(Z,
θ
, V, T)
Response time + Contact checking time × 2 + (Self calibration time) + V sampling time +
V calculation time + Measurement circuit switching time+ Sample delay time + Z sampling
time + (Measurement signal zero cross detection time) + Z calculation time
Function (R, X, T)/(Z,
θ
, T)
Response time + Contact checking time × 2 + Measurement circuit switching time +
Sample delay time + Z sampling time + (Measurement signal zero cross detection time) +
Z calculation time
Function (V, T)
Response time + Contact checking time × 2 + (Self calibration time) + V sampling time +
V calculation time
(Self calibration time is added when the self calibration is set to [AUTO].)
(Measurement signal zero cross detection time is added when the measurement signal
zero cross stop function is [ON].)
10.2 Additional Function
Measurement
Range setting
Function overview Setting measurement range of impedance. (Voltage and temperature have no setting due
to the single range.)
Function setting 3 m
Ω
/10 m
Ω
/100 m
Ω
(AUTO setting is not provided.)
Setting backup Yes
Frequency setting
Function overview Setting the measurement frequency of impedance measurement.
Function setting 0.10 Hz to 1050 Hz
Setting backup Yes
Measurement speed setting
Function overview Setting impedance measurement, Setting measurement speed of voltage measurement.
Function setting Impedance measurement FAST/MED/SLOW
Voltage measurement FAST/MED/SLOW
Setting backup Yes
Function setting
Function overview Setting measurement functions.
Function setting
(R, X, V, T)/(Z,
θ
, V, T)/(R, X, T)/(Z,
θ
, T)/(V, T)
Setting backup Yes
Setting trigger source
Function overview Trigger for measurement start.
Function setting EXT/INT
EXT: External trigger
INT: Internal trigger
(The voltage limit is turned ON when the internal trigger is set.)