102
Specications of Measurement Functions
Measuring current/DC load (DC load is the offset current that is applied to the measuring object when measuring
the impedance.)
3 m
Ω
range 10 m
Ω
range 100 m
Ω
range
Measurement current 1.5 A rms ±10% 500 mA rms ±10% 50 mA rms ±10%
DC load current 1 mA or less 0.35 mA or less 0.035 mA or less
Measurement wave
number
FAST MED SLOW
0.10 Hz to 66 Hz 1 wave 2 waves 8 waves
67 Hz to 250 Hz 2 waves 8 waves 32 waves
260 Hz to 1050 Hz 8 waves 32 waves 128 waves
Overrange indication OverRange
Voltage measurement
Measurement terminal
structure
BNC
Measurement terminal
function
SENSE-H terminal Voltage detection terminal
SENSE-L terminal Voltage detection terminal
Measurement items Voltage (Parameter indication: V)
Range structure 5 V (single range)
Display range -5.10000 V to 5.10000 V
Resolution 10 µV
Measurement speed
setting
FAST/MED/SLOW
Measurement time FAST 0.1 s
MED 0.4 s
SLOW 1.0 s
(When self calibration is AUTO, 210 ms is added to the measurement time.)
Sampling period 6 kHz
Overrange display OVER VOLTAGE
Temperature measurement
Measurement terminal
structure
Four-terminal earphone jack
φ
3.5 mm
Measurement items Temperature (Parameter indication: T)
Display range -10.0°C to 60.0°C
Resolution 0.1°C
Sampling time 2.3 s
Overrange indication +Over°C, -Under°C
Indication when
unconnected
--.-°C