EasyManua.ls Logo

Hioki BT4560 - Page 119

Hioki BT4560
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
113
Accuracy
Accuracy graph
3 m
Ω
range (0.1 Hz to 100 Hz), 10 m
Ω
range, and 100 m
Ω
range
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
0.0
-180 -90 0 90 180
X R
Accuracy [%rdg.]
Phase [°]
Impedance accuracy excluding α (0.004|R| + 0.0017|X|, 0.004|X| + 0.0017|R|)
3 m
Ω
range (110 Hz to 1050 Hz)
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
0.0
-180 -90 0 90 180
X R
Accuracy [%rdg.]
Phase [°]
Impedance accuracy excluding α (0.004|R| +0.0052|X|, 0.004|X| +0.0052|R|)
Voltage measurement accuracy
Voltage measurement
V
Display range -5.10000 V to 5.10000 V
Resolution 10 µV
Voltage accuracy
FAST ±0.0035% rdg.±5 dgt.
MED ±0.0035% rdg.±5 dgt.
SLOW ±0.0035% rdg.±5 dgt.
Temperature
coefcient
±0.0005% rdg.±1 dgt./°C
(Applied in the ranges of 0°C to 18°C, and 28°C to 40°C)
10
Specications

Table of Contents

Related product manuals