113
Accuracy
Accuracy graph
• 3 m
Ω
range (0.1 Hz to 100 Hz), 10 m
Ω
range, and 100 m
Ω
range
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
0.0
-180 -90 0 90 180
X R
Accuracy [%rdg.]
Phase [°]
Impedance accuracy excluding α (0.004|R| + 0.0017|X|, 0.004|X| + 0.0017|R|)
• 3 m
Ω
range (110 Hz to 1050 Hz)
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
0.0
-180 -90 0 90 180
X R
Accuracy [%rdg.]
Phase [°]
Impedance accuracy excluding α (0.004|R| +0.0052|X|, 0.004|X| +0.0052|R|)
Voltage measurement accuracy
Voltage measurement
V
Display range -5.10000 V to 5.10000 V
Resolution 10 µV
Voltage accuracy
FAST ±0.0035% rdg.±5 dgt.
MED ±0.0035% rdg.±5 dgt.
SLOW ±0.0035% rdg.±5 dgt.
Temperature
coefcient
±0.0005% rdg.±1 dgt./°C
(Applied in the ranges of 0°C to 18°C, and 28°C to 40°C)