47
Prevents Charging and Discharging due to the Measurement Signal (Measurement Signal Zero Cross Stop Function)
4.7 Prevents Charging and Discharging due to the
Measurement Signal (Measurement Signal Zero
Cross Stop Function)
This function performs the process of stopping the applied measurement signal at zero cross during
impedance measurement to prevent charging and discharging of the object to be measured. When
the measurement signal zero cross stop function is enabled, the measurement time increases by
approximately one cycle of measurement frequency.
I
0
Charging
Discharging
The charging and discharging of the battery is prevented
by stopping the measurement current at zero cross.
1
Press (MENU). (Settings screen is displayed.)
2
Select [MEAS] tab.
Selection
4
Customization of Measurement Conditions

Table of Contents