EasyManua.ls Logo

Hioki BT4560 - Page 98

Hioki BT4560
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92
Internal Circuitry
PNP setting
PLC, others
Output
Common
Do not connect external power supply to 8 pin.
The instrument
8
ISO_5V
1
START (TRIG)
2
0ADJ_ALL
10
ERR
11
RorZ_HI
9
ISO_COM
27
ISO_COM
PLC, others
Input
Common
2 k
Ω
1 k
Ω
10
Ω
Zener voltage 30 V
Internally isolated common
(This is isolated to the protective ground.)
PNP
EXT.I/O MODE
selector
Internally isolated
power supply
Share the ISO_COM for the common terminals of the input and the output signal.
Electrical Specications
Input signal Input type Photo-coupler-isolated, non-voltage contact inputs (corresponding
to current sink/source output)
Input ON Residual voltage 1 V (Input ON Current 4 mA (reference value))
Input OFF OPEN (Breaking current less than 100 µA)
Output signal Output type Photo-coupler-isolated open drain output (non-polarity)
Maximum load voltage 30 V max DC
Maximum output
current
50 mA/ch
Residual voltage Less than 1 V (Load current 50 mA)/less than 0.5 V (Load current
10 mA)
Internally
isolated power
supply
Output voltage Corresponding to sink output: +5.0 V±10%, Corresponding to
source output: -5.0 V±10%
Maximum output
current
100 mA
External power input None
Insulation Floating from the protective grounding potential and the
measurement circuit
Insulation rating Voltage to ground 50 V DC, 33 V AC rms, less than 46.7 Vpeak
AC

Table of Contents

Related product manuals