101
Setting the TEST Signal OFF Timing
7.4 Setting the TEST Signal OFF Timing
There are two timing options for the EXT.I/O TEST signal output to return from LOW
to HIGH when a test is completed.
1
Open the Settings screen.
Select [SET].
2
Select the operation mode for [EXT.I/O TEST-PIN].
Selection
[SLOW]
[FAST]
Return to the
Measurement
screen.
[SLOW]
After a test is completed, LOW (status during the test)
is maintained until the voltage of the equipment to
be measured drops to approximately 10 V due to the
discharge function.
[FAST]
Regardless of the level of discharge, the status returns to
HIGH almost simultaneously after a test is completed.
10
9
8
7
6
5
4
3
2
1
Appx. Ind.