Setting the Test Duration and Response Time
Setting the response time
Response time is the time to prohibit comparator judgment after a test starts until
the set response time elapses. No measured value is displayed during the response
time. The response time is included in the test duration. The same response time is
applicable at any test voltage.
Test starts Comparator judgment starts
Response
time
Measurement value display,
comparator judgment
Test duration
The response time varies depending on the equipment to be measured.
If a capacitive load is connected, it may take time for the output voltage to
rise.
When the response time is set to [AUTO], the output is monitored and the
measurement starts after the voltage is stabilized. The time required for the
voltage to stabilize varies depending on the equipment to be measured.
Iftheelectrostaticcapacityoftheequipmenttobemeasuredis1µFor
more, the [AUTO] function may not operate properly because it takes time
to charge the equipment to be measured. Check the time required for the
value on the voltage monitor to reach the set voltage, and then set it after
[MANUAL] is selected.
When the response time is set to [MANUAL], the measurement starts
anytime. The measurement may start before the output voltage rises
depending on the equipment to be measured. Set the response time
according to the equipment to be measured.
During the response time, [D] is displayed to the left of the test duration
indication on the measurement screen.