49
4
Testing
This chapter describes testing procedures for the instrument.
4.1 Starting Measurements
WARNING
Observe the following to avoid electric shock.
When the TEST lamp is blinking, do not touch the equipment to be
measured, the tip of the probe, and measurement terminals.
The measurement terminals on the front panel and the
corresponding ones on the rear panel are connected each other in
the instrument. Do not touch the measurement terminals.
Do not touch the equipment to be measured, the tip of the probes,
and measurement terminals immediately after tests are completed.
High voltage charge may cause electric shock.
To avoid electric shock, use the discharge function on the
instrument to release charge in pieces of equipmentS to be
measured after tests. (p. 54)
1
Set the measurement items. (p. 33)
2
Press the START key to start a measurement.
START key
3
The TEST lamp blinks at intervals of 250 ms during measurement.
TEST lamp
4 Testing
10
9
8
7
6
5
4
3
2
1