49
4
Testing
This chapter describes testing procedures for the instrument.
4.1 Starting Measurements
WARNING
Observe the following to avoid electric shock.
• When the TEST lamp is blinking, do not touch the equipment to be
measured, the tip of the probe, and measurement terminals.
• The measurement terminals on the front panel and the
corresponding ones on the rear panel are connected each other in
the instrument. Do not touch the measurement terminals.
• Do not touch the equipment to be measured, the tip of the probes,
and measurement terminals immediately after tests are completed.
High voltage charge may cause electric shock.
• To avoid electric shock, use the discharge function on the
instrument to release charge in pieces of equipmentS to be
measured after tests. (p. 54)
1
Set the measurement items. (p. 33)
2
Press the START key to start a measurement.
START key
3
The TEST lamp blinks at intervals of 250 ms during measurement.
TEST lamp
4 Testing
10
9
8
7
6
5
4
3
2
1
Appx. Ind.