19
Measurement Workow
1.3 Measurement Workow
Be sure to refer to “Usage Notes” (p. 7) before use.
1
Preparing for measurement (p. 23)
(Use the measurement terminals on the rear
panel in case of 4-terminal measurements.)
2
Checking before measurement (p. 27)
3
Setting measurement conditions (p. 33)
Item Description Reference
Test voltage
Select from 25 V to 1000 V.
p. 34
Resistance range
Select either manual range (2 M
Ω
, 20 M
Ω
, 200 M
Ω
,
2000 M
Ω
, 4000 M
Ω
) or auto range.
p. 36
Measurement speed
Select either FAST or SLOW.
p. 38
Test duration
Set how long to apply the test voltage (test duration
function).
p. 39
Set the response time.
p. 41
Comparator
(upper/lower limit values)
Set the upper/lower resistance values.
p. 44
Test mode
Select from continuous mode, FAIL STOP mode, PASS
STOP mode, or forced termination judgment mode.
p. 46
Beep sound
Set the conditions to announce the judgment result
with a beep sound.
p. 47
4
Performing measurement
5
Turning off the power after use
10
9
8
7
6
5
4
3
2
1
Appx. Ind.