155
Test duration
Denition of test
duration
Test duration = Response time + Measurement time
Setting range
Setting range 0.045 s (45 ms) to 999.999 s
Test duration ON/OFF (- -)
Display 000.000 s setting
Screen display: 000.0 s, 0.000 s if less than 100 ms
“000.0 s” is displayed when OFF is set, “HOLD” is displayed at the end
of a test.
Operation The test duration counts down from the specied time after a test
starts.
When OFF is set, the test duration counts up after a test starts.
“999.9 s” is displayed when the test duration exceeds the display
range.
Guaranteed
accuracy range
0.050 s (50 ms) to 999.999 s
Set resolution 1 ms
Test duration
accuracy
±0.005 s (setting range 0.045 s to 0.099 s)
±0.05 s (setting range 0.100 s to 9.999 s)
±0.5 s (setting range 10.000 s to 999.999 s)
Voltage (V)
Set voltage
Response time
Charge time
Measurement time
Discharge time
Test duration
Time (t)
FAST
SLOW
Time diagram
(FAST, SLOW: TEST signal OFF timing setting)
10
9
8
7
6
5
4
3
2
1