71
Setting the Frequency of Power Supply Manually
5.8 Setting the Frequency of Power Supply
Manually
There are three power supply frequency settings: AUTO, 50 Hz, and 60 Hz.
To eliminate any noise, the supply power frequency needs to be set properly.
In the default setting, the power supply frequency is supposed to be identied
automatically (AUTO) when the power is supplied, however, it also can be set
manually.
If the power supply frequency is not set properly, a measured value does not stabilize.
Even in the AUTO setting, if the power supply frequency cannot be detected properly
due to signicant power noise, an error (ERR:097) is displayed when the power is
supplied. Set the frequency according to the power supply.
1
Open the Settings screen.
Select [SET].
2
Set [POWER FREQ].
(When [AUTO]isset,theautomaticallyidentiedpowerfrequencyis
displayed on the right.)
Selection
[AUTO]
[50Hz]
[60Hz]
Return to the
Measurement
screen.
The power frequency is identied automatically only once when the power is
supplied.
After the power supply frequency setting is changed from [50Hz] or [60Hz] to
[AUTO], turn off the power supply and then turn it on again.