Switching the Measurement Speed (FAST/SLOW)
3.3 Switching the Measurement Speed (FAST/
SLOW)
There are two measurement speeds, [FAST] and [SLOW]. The resistance is
measured and displayed in 30 ms when the measurement speed is set to [FAST] and
in 500 ms when the speed is set to [SLOW].
Setting the measurement speed to [SLOW] is effective if a measured value uctuates
and it is difcult to read the value.
1
Open the Settings screen.
Select [SET].
2
Set the test speed.
Selection
[SLOW]
[FAST]
Return to the
Measurement
screen.
IMPORTANT
• The sampling time is 100 ms even when the measurement speed is set to [FAST]
if the contact check function is ON.
(The sampling time is 500 ms if the measurement speed is set to [SLOW]
regardless of the function check setting.)
• If the test duration is set to less than 500 ms when the measurement speed is set
to [SLOW], the measured value is not displayed. Set a test duration that is longer
than the sampling time.