156
Response time
Denition of
response time
No-judgment time
Response time function
Setting range AUTO, 0.005 s (5 ms) to 999.999 s
Operation Comparator judgment is not performed after measurement starts until
the response time elapses. The timer counts down along with the test
duration timer (Test voltage is generated).
Resolution 1 ms
Functions
Voltage monitoring
Update speed Same as that of resistance display
Contact check
Method 4-terminal (For 2-terminal, the comparator upper and lower limit values
are used.)
Operation Contact check is performed for each sampling and a contact error
occurs if a faulty contact is detected. (Obtained by using RS command
or EXT.I/O)
Sampling Sampling varies depending on the measurement speed setting.
FAST: 100 ms, SLOW: 500 ms
Comparator
Display The upper and lower limit values are always displayed.
Judgment UPPER_FAIL, PASS, LOWER_FAIL, UL_FAIL
UPPER_FAIL Measured value ≥ Upper limit value
PASS
Upper limit value > Measured value > Lower limit value
LOWER_FAIL Measured value ≤ Lower limit value
UL_FAIL When no judgment can be made.
Beep sound ON/OFF (based on the judgment beep sound setting)
Display The LEDs for PASS, UPPER_FAIL, LOWER_FAIL are lit.
Both UPPER_FAIL and LOWER_FAIL are lit at the same time for UL_
FAIL.
“----” is displayed when the comparator is set to OFF.
Judgment output Based on the test mode.
Both UPPER_FAIL and LOWER_FAIL are output at the same time for
UL_FAIL.
The last test result is held.