Operation No judgment is performed for forced termination (STOP key, STOP
signal).
However, a judgement result is output when forced termination
judgment mode is set.
(No judgment is performed if the STOP key is pressed before one
measurement, with an elapsed time of less than 45 ms, is not
completed.)
No LED display or EXT.I/O signal output.
Memory function
Items to be
saved
Rated measurement voltage, comparator upper and lower limit values
(including comparator ON/OFF), test mode, judgment beep sound,
test duration, response time (no-judgment time), resistance range (ON/
OFF of auto range, including resistance range for xed range), and
measurement speed
Number of
memory data
units
Maximum of 10 patterns (can be saved and loaded)
Measurement speed
FAST Sampling: 30 ms/time
SLOW Sampling: 500 ms/time
Data hold Automatically holds the measurement value display at the time of
measurement completion, judgment result display, EXT.I/O judgment
result output, and analog output.
10
9
8
7
6
5
4
3
2
1