EasyManua.ls Logo

Hioki ST5520-01 - Page 166

Hioki ST5520-01
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
158
Test mode Switch a mode selecting from continuous mode, PASS STOP mode,
FAIL STOP mode, or forced termination judgment mode.
Continuous mode
Measurement is performed for the specied test duration, judgment
is performed for each measurement, the judgment result is output,
and the output voltage is cut off.
The displayed measured value and judgment result display are held
as they are at the end of a test.
PASS STOP mode
A test is completed when a PASS judgment is made and the output
voltage is cut off.
The displayed measured value and PASS display are held as they
are at the end of the test.
FAIL STOP mode
A test is completed when a FAIL judgment is made and the output
voltage is cut off.
The displayed measured value and FAIL display are held as they are
at the end of the test.
Forced termination judgment mode
The judgment result based on the upper and lower limit values is
displayed after the setting time elapsed or at the time of STOP is
input (key, RS command, or EXT.I/O).
Judgment result is output on the display, via RS command, or via
EXT.I/O.
The displayed measured value and judgment result display are held
as they are at the end of a test.
Double action
Operation The START key is enabled only for 1 s after the STOP key is pressed.
Key lock function
Reset means Hold down the [UNLOCK] key (for 1 s) when the key lock function is
enabled.
Display [LOCK] is displayed at the top of the screen when the key lock
function is enabled.
(The START key and STOP key are enabled.)
Interlock function
Display [LOCK] is displayed at the top of the screen when the interlock
function is enabled.
Operation No key operations are enabled.
No tests can be started via EXT.I/O or RS command.

Table of Contents

Related product manuals