126
Command Reference
8.6 Command Reference
Command reference description
Syntax
:MOHM:RANGe <data>
Reference
Species the syntax for the command.
<>:
Species the description of the message data (text or numerical
parameter). A response is returned in capital letters for a text parameter.
Numerical parameter:
NRf: Format including NR1, NR2, and NR3.
NR1: Integer data (example: +12, -23, 34)
NR2: Decimal data (example: +1.23, -23.45, 3.456)
NR3: Floating decimal point index data (example: +1.0E-2, -2.3E+4)
<data> 2 M / 20 M / 200 M / 2000 M / 4000 M / AUTO (text data)
Reference
Explains parameter data.
Example Set to 200 M
Ω
range.
PC>
:MOHM:RANGe 200M
Reference
Example of the use of the command
PC> indicates a command from the computer.
ST5520> indicates a response from the ST5520.
Error An execution error occurs if the 2000 M
Ω
range is executed when the test
voltage is between 25 V and 99 V.
Reference
Explains the error.
Program message