105
Using Analog Output
Connecting the output cord
The analog output terminal on the rear panel outputs direct voltage proportional to the
resistance.
When connecting to a recorder, etc., use an output cord with an input resistance of 1
M
Ω
or more.
When the input resistance is low, accurate measurement cannot be performed.
CAUTION
To avoid damage to the instrument or the output cord, do not input voltage
or current externally to the analog output terminal. Do not connect the
output cord plug to the measurement terminal of the instrument.
Setting the analog output
1
Open the Settings screen.
Select [SET].
2
Select [ANALOG OUT RANGE].
Selection
[FULL]
[EACH]
Return to the
Measurement
screen.
[FULL]
0 to 4 V is output within the all measured resistance ranges.
[EACH]
The voltage is output according to each resistance range.