EasyManua.ls Logo

Hioki ST5520-01 - Input; Guaranteed Accuracy

Hioki ST5520-01
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
154
Resistance
measurement
accuracy
Commonly Applicable to FAST and SLOW
Rated measurement
voltage
Guaranteed
accuracy range
Basic accuracy
25 V ≤ V < 100 V
0.000 M
Ω
to 2.000 M
Ω
±2% rdg. ±5 dgt.
1.90 M
Ω
to 20.00 M
Ω
19.0 M
Ω
to 200.0 M
Ω
±5% rdg.
100 V ≤ V < 500 V
0.000 M
Ω
to 2.000 M
Ω
±2% rdg. ±5 dgt.
1.90 M
Ω
to 20.00 M
Ω
19.0 M
Ω
to 2000 M
Ω
±5% rdg.
500 V ≤ V1000 V
0.000 M
Ω
to 2.000 M
Ω
±2% rdg. ±5 dgt.1.90 M
Ω
to 20.00 M
Ω
19.0 M
Ω
to 200.0 M
Ω
190 M
Ω
to 4000 M
Ω
±5% rdg.
4010 M
Ω
to 9990 M
Ω
±25% rdg.
(The temperature coefcient is based on “Guaranteed
accuracy” (p. 154).)
Input
Over voltage input
protection
1100 V DC (positive polarity only)
Guaranteed accuracy
Conditions of
guaranteed
accuracy
Warming-up time 15 minutes or more
Temperature
and humidity
for guaranteed
accuracy
23°C ± 5°C (73°F ± 9°F), 80% RH or less, no condensation
Temperature
coefcient
Add 0.1 × Basic accuracy × (T-23) T: Operating temperature (°C)
Accuracy warranty
period
1 year