108
Using the Switched Probe
7.8 Using the Switched Probe
The optional switched probe enables the start and stop operations of the ST5520/
ST5520-01 while the probe is held in hand.
Before a test, set the push switch mode to either trigger ([TRIG]) or continuous
([CONT]). It is initially set to continuous mode ([CONT]).
1
Open the Settings screen.
Select [SET].
2
Select the operation mode for [EXT.SWITCH MODE].
Selection
[CONT]
[TRIG]
Return to the
Measurement
screen.
[CONT]
(Continuous mode)
A test is performed only when the push switch is pressed.
When the push switch is released, the test is terminated.
[TRG]
(Trigger mode)
Press the push switch to start a test. The test continues
even when the push switch is released. When the push
switch is pressed during the test, the test is terminated.
For the operation of the ST5520 and ST5520-01 when the switched probe is
used, refer to “Enabled/disabled input list” (p. 163).
The 9299 Switched Probe can be used as an ordinary probe when the switch
signal line plug is not connected to the EXT.SW terminal.