Loading...
152
Charge (measurement)
current
1.8 mA
Short circuit current Less than 2.0 mA
Analog output Not isolated against the power supply protective ground terminal
and isolated against the measurement terminals.
Output voltage: Switching between FULL and EACH
1. FULL 0 to 4 V is output in all measured resistance ranges.
Measurement voltage Displayed resistance
Output voltage
(DC)
25 V ≤ V < 100 V 0.000 M
Ω
to 200.0 M
Ω
0 V to 4 V
100 V ≤ V < 500 V 0.000 M
Ω
to 2000 M
Ω
0 V to 4 V
500 V ≤ V ≤ 1000 V 0.000 M
Ω
to 4000 M
Ω
0 V to 4 V
All measurement
voltages
Over. F 4 V
Under. F 0 V
2. EACH 0 to 4 V is output according to each resistance range.
Resistance range Displayed resistance
Output voltage
(DC)
2 M
Ω
0.000 M
Ω
to 2.000 M
Ω
0 V to 4 V
20 M
Ω
1.90 M
Ω
to 20.00 M
Ω
0.38 V to 4 V
200 M
Ω
19.0 M
Ω
to 200.0 M
Ω
0.38 V to 4 V
2000 M
Ω
(100 V ≤ V < 500 V)
190 M
Ω
to 2000 M
Ω
0.38 V to 4 V
4000 M
Ω
(500 V ≤ V ≤ 1000 V)
190 M
Ω
to 4000 M
Ω
0.19 V to 4 V
All resistance ranges
Over. F 4 V
Under. F 0 V
Analog output
accuracy
±2% f.s.
Load resistance 1 M
Ω
or more

Table of Contents

Related product manuals