Troubleshooting
Display Meaning Solution Reference
ERR:004 Measurement
aborted by
interlock.
Interlock error The test is canceled because the
interlock function is activated during
the test.
p. 106
ERR:030 Command
error.
Command error Check that the command is correct.
-
ERR:031 Execution
error.
(Parameter
error)
Execution error.
The parameter
setting is outside
of the range.
Check that the parameter range is
correct.
-
ERR:032 Execution
error.
Execution error Check that the command with
an error does not fall under the
execution error conditions.
-
ERR:090 ROM check
sum error.
Program ROM
check sum error
The instrument is malfunctioning.
Send the instrument out for repair.
-
ERR:091 RAM error. CPU RAM error The instrument is malfunctioning.
Send the instrument out for repair.
-
ERR:094 Output
voltage error.
Power off by
rear power-
sw.
Output voltage
error.
Turn off the
power switch on
the rear panel.
A voltage that is higher than the set
test voltage may be output.
Refer to “Troubleshooting” as well. p. 167
ERR:096 Backup data
error.
Setting backup
error
The settings have been initialized.
Reset the measurement conditions,
etc.
-
ERR:097 Power line
detection
error.
Select power
line cycle.
Power frequency
detection error
Set the frequency according to the
supply power.
p. 71