35
Setting the Test Voltage
Setting the test voltage in the Settings screen
1
Open the Settings screen.
Select [SET].
2
Set [VOLTAGE] to edit mode.
Selection
Edit the value.
Return to the
Measurement
screen.
3
Changeandthenconrmthetestvoltagevalue.
Change the value.
(Up/Down keys)
Change the cursor
position. (Left and
right keys)
Conrm the value.
Cancel the change.
When a voltage that is higher than the set test voltage is output (set voltage +
20%), [Output voltage Error] is displayed on the screen. The test is stopped and
no further key operations can be performed. Turn OFF the power switch on the rear
panel.
Refer to “10.2 Troubleshooting” (p. 167).