Timing Chart
Continuous test mode timing chart (2)
When the test duration setting (TIMER) is ON and the measurement is performed by the /START
signal input from EXT.I/O or when the test duration setting (TIMER) is ON and the measurement
is performed by the START key on the panel
Setting
Test duration (TIMER) Response time (DELAY) TEST signal OFF timing
ON, 45 ms to 999.9 s AUTO, 5 ms to 999.9 s FAST
Measurement stop
Measurement-1 Measurement-2
Previous measured value
Previous result
Previous result
Previous result
0/0.0/0.00 M
Ω
Measured value-1 Measured value-2
Measured value-2 result
Measured value-2 result
Measurement-1 contact check result
Measurement,
judgment
(including
contact check)
10 V
OFF
OFF
OFF
/START
(Input)
/TEST
(Output)
Test voltage
Measured
value display
/PASS
(Output)
/U.FAIL
/L.FAIL
(Output)
/ERR
(Output)
t1
t2
t4
t3
t6
t5
t7
t5
When the TEST
signal OFF
timing setting is
SLOW