5 - 16
Transpector MPH Operating Manual
NOTE: ALARM/WARNING SETTINGS and Transpector MPH Scan Parameters
each have a SET button next to the changeable parameters.
The SET buttons are not interchangeable.
To change alarm and warning settings, click SET next to the ALARM
and WARNING group.
To change scan parameters, click SET next to the MASS and DWELL
group.
3 Thermometer BarThe Thermometer bar shows the real-time level of the
signal. The bar will raise and lower based on the signal in amps of the specified
mass. If the signal exceeds warning or alarm levels, the color of the bar will
change. See Figure 5-13.
Figure 5-13 Thermometer bar levels
4 EMISSION and EM ControlThese work the same as EMISSION and EM in
the MONITOR window (section 5.5.2.1.1, Editable Parameters, on page 5-11).
Click EMISSION or EM to turn on or off the emission or EM.
5 Mass Trend GraphThe mass trend graph will display how the signal
(in amps) for the specified mass changes with time. There are two lines
(one yellow and one red) on the trend graph. These lines represent warning
and alarm levels. When the signal increases above the alarm level lines, a leak
is present.