INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 103 / 152
Type Notification Possible sources of
error
Remedy
E340 Emission error Pressure too high with
aged cathode
Contact customer service
W342 Cathodes not connected Both cathodes
defective during self-
test
Contact customer service
Plug of ion source not
connected
Contact customer service
W350 Suppressor not connected Suppressor cable not
connected or defective
Contact customer service
E352 Preamplifier not connected Preamplifier defective
or cable not connected
Contact customer service
W358 Alternate preamplifier
ranges
Preamplifier defective Contact customer service
Unusual leak rate
waveform
Contact customer service
W359 Overdriven preamplifier Ion source or
preamplifier defective
Contact customer service
Mass spectrometer
contaminated
Contact customer service
W360 Preamplifier output signal
too low
Ion source defective Contact customer service
Mass spectrometer
contaminated
Contact customer service
W361 The offset voltage of the
preamplifier is too high
Preamplifier defective Contact customer service
W362 Preamplifier range error Preamplifier defective Contact customer service
MSB defective Contact customer service
W390 500 GOhm amplifier resistor
out of range
Preamplifier defective Contact customer service
MSB defective Contact customer service
E400 TMP error message Unhandled error
message from the TMP
Contact customer service
W401 TMP warning message Unhandled warning
message from the TMP
Contact customer service
E402 No communication with
turbo pump converter
TMP cable defective or
not connected
Contact customer service
TMP converter
defective
Contact customer service
MSB defective Contact customer service