INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 105 / 152
Type Notification Possible sources of
error
Remedy
E451 Temperature prewarning of
backing pump
The ambient
temperature is too high
Reduce the temperature in the
environment where the device is located
Ambient temperature is
too low
Increase the temperature in the
environment where the device is located
Fan blocked or
defective
Contact customer service
Temperature sensor of
backing pump is
defective
Contact customer service
W452 No communication with
backing pump
Backing pump control
cable defective or not
connected
Contact customer service
Backing pump has no
power supply
Contact customer service
Backing pump
converter defective
Contact customer service
VI-Board or MSB
defective
Contact customer service
E453 Backing pump failure Unhandled error
message from the
backing pump
Contact customer service
E454 Fault in backing pump:
Undervoltage
The backing pump in
the leak detector has
insufficient mains input
voltage
Make sure that you use the supplied
power cord
Have your mains voltage installation
checked by a qualified electrician.
Backing pump
converter defective
Contact customer service
W455 Backing pump rotation
speed too low
No run-up of backing
pump
Contact customer service
W456 Operation of the backing
pump is locked
The plug at connection
P1 of the frequency
converter of the
backing pump is
defective or not
connected
Contact customer service