7 | Warning and error messages INFICON
106 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
Type Notification Possible sources of
error
Remedy
E457 Fault in backing pump:
Power failure
The backing pump in
the leak detector has a
faulty mains input
voltage
Make sure that you use the supplied
power cord
Visually inspect the power cord for
damage
Have your mains voltage installation
checked by a qualified electrician.
W458 The backing pump is not
started because it is too
cold. Delete the warning at
your own risk to start the
pump.
Ambient temperature is
too low
Increase the temperature in the
environment where the device is located
If you bring the leak detector inside from a
cold outside area, wait for at least an hour
before you switch on the leak detector
Temperature sensor of
backing pump is
defective
Contact customer service
Backing pump
converter defective
Contact customer service
W459 Run-up of backing pump
takes too long
Exhaust connection
sealed
Make sure that the exhaust hose is
correctly connected
Backing pump
defective
Contact customer service
W460 No communication with the
TMP2 converter
TMP2 cable defective
or not connected
Contact customer service
TMP2 converter
defective
Contact customer service
MSB defective Contact customer service
E461 Unhandled error message of
TMP2
TMP2 converter
defective
Contact customer service
TMP2 defective Contact customer service
E462 Temperature of TMP2 pump
too high
The ambient
temperature is too high
Reduce the temperature in the
environment where the device is located
Fan blocked or
defective
Contact customer service
TMP2 defective Contact customer service