INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 107 / 152
Type Notification Possible sources of
error
Remedy
E463 Temperature error of
backing pump
The ambient
temperature is too high
Reduce the temperature in the
environment where the device is located
Fan blocked or
defective
Contact customer service
Ambient temperature is
too low
Increase the temperature in the
environment where the device is located
Temperature sensor of
backing pump is
defective
Contact customer service
E464 Power consumption of
backing pump too high
Exhaust connection
sealed
Make sure that the exhaust hose is
correctly connected
Heavy duty operation Contact customer service
E465 Incorrect flow direction in
backing pump
Gas flow into device
due to pressure at
exhaust connection
Remove the pressure from the exhaust
connection
E466 Run-up error of backing
pump / rotation speed too
low
Exhaust connection
sealed
Make sure that the exhaust hose is
correctly connected
Backing pump
defective
Contact customer service
E467 Overload protection of
backing pump
Exhaust connection
sealed
Make sure that the exhaust hose is
correctly connected
Backing pump
defective
Contact customer service
W468 Unhandled warning
message of backing pump
Unhandled error
message from the
backing pump
Contact customer service
E500 Pressure sensor p1 not
connected
Pressure sensor not
connected or cable
defective
Contact customer service
VI-Board or MSB
defective
Contact customer service
E502 Pressure sensor p2 not
connected
Pressure sensor not
connected or cable
defective
Contact customer service
VI-Board or MSB
defective
Contact customer service