23
PROBLEM
OTHER
SYMPTOMS
POSSIBLE
CAUSE
SOLUTION
Documentation
Bad connections
Check all cabling.
See chapter 8.3.5.
Fuse burnt out
Check cabling and
replace fuse
See chapters 8.3.5 and
8.3.4
Batteries
defective
Replace batteries
See chapter 8.3.2.
Cabling to power
module or
remote defective
check cabling
See chapter 8.3.5.
None
Power module
defective
Replace power
module
See chapter Fehler!
Verweisquelle konnte
nicht gefunden werden..
Motors do
not run
Corroded
contacts
Water, salt or
urine has
penetrated
Check cabling,
replace if
necessary.
See chapter 8.3.5.
Bearing
defective
Replacing the
motor
See chapter 8.17.1.
Motor
makes
clicking
noise
None
Collector
defective
Measure internal
resistance of
motor, replace
motor if defective.
See chapters 7.1 and
8.17.1
Bearing
defective
Replacing the
motor
See chapter 8.17.1.
Scraping
noise or
motor
blocked
None
Transmission
defective
Replace
transmission.
See chapter 8.17.1.
Transmission
defective
Replace
transmission.
See chapter 8.17.1.
Transmissi
on makes
clicking
noise
None
Drive wheel
loose
Tighten drive
wheel, secure
bolts with Loctite if
necessary
See chapter 8.16.1.

Table of Contents

Other manuals for Invacare STORM

Related product manuals