STATus Commands
Copyright © ITECH Electronic Co., Ltd. 120
Query syntax
STATus:OPERation[:EVENt]?
Arguments
None
Returns
<NR1>
Related syntax
STATus:OPERation:ENABle
Bit definition of operation event register:
Bit
7
6
5
4
3
2
1
0
Name
n.u
n.u
n.u
n.u
n.u
WTG
LIST
CAL
Value Bit
Weight
4
2
1
STATus:OPERation:CONDition?
This command can read the parameter from the operation condition register.
This is a read-only register, which directly reflects those bits in the operating state
of the instrument are set to 1 or 0, and will not be latched. After the command is
executed, the value of the query condition register will not be cleared.
Query syntax
STATus:OPERation:CONDition?
Arguments
None
Example
STATus:OPERation:CONDition?
Returns
<NR1> This is a decimal value with a binary weighted sum.
STATus:OPERation:ENABle <NR1>
Sets the value of the enable register for the Operation Status group. The enable
register is a mask for enabling specific bits from the Operation Event register to
set the OPER (operation summary) bit of the Status Byte register.
STATus:PRESet clears all bits in the enable register. *CLS does not clear the
enable register, but does clear the event register.
Syntax
STATus:OPERation:ENABle <NR1>