FETCh & MEASure Commands
Copyright © ITECH Electronic Co., Ltd. 60
FETCh[:SCALar]:CURRent:AMPlitude:MAXimum:POSitive?
MEASure[:SCALar]:CURRent:AMPlitude:MAXimum:POSitive?
This command is used to read the positive peak value of single-phase current.
Note: this command parameter is optional. If there is no parameter, it will return
the value of A, B, C three-phase by default.
Syntax:
FETCh[:SCALar]:CURRent:AMPlitude:MAXimum:POSitive?
[A|B|C|CH1|CH2|CH3]
Arguments:
[A|B|C|CH1|CH2|CH3]
Returns:
<NRf>…
Example:
FETC:CURR:AMP:MAX:POS?
FETCh[:SCALar]:CURRent:AMPlitude:MAXimum:NEGative?
MEASure[:SCALar]:CURRent:AMPlitude:MAXimum:NEGative?
This command is used to read the negative peak value of single-phase current.
Note: this command parameter is optional. If there is no parameter, it will return
the value of A, B, C three-phase by default.
Syntax:
FETCh[:SCALar]:CURRent:AMPlitude:MAXimum:NEGative?
[A|B|C|CH1|CH2|CH3]
Arguments:
[A|B|C|CH1|CH2|CH3]
Returns:
<NRf>…
Example:
FETC:CURR:AMP:MAX:NEG?