SOURce Commands
Copyright © ITECH Electronic Co., Ltd. 40
Query syntax:
FREQuency[:IMMediate]? [MAXimum|MINimum| DEFault]
Returns:
MAXimum|MINimum|DEFault
*RST:
50Hz
Example
FREQuency 300.0
FREQuency?
[SOURce:]PFREquency[:IMMediate] <phase>,<NRf+>
Set and query the frequency value of a phase.
Syntax:
PFREquency[:IMMediate] <phase>,<NRf+>
Query syntax:
PFREquency[:IMMediate]? <phase>
Example
PFREQuency A,300.0 //Set the frequency of phase A to 300.0Hz
PFREQuency? A //Query the frequency value of phase A
[SOURce:]FREQuency:SLOPe[:IMMediate]
<NRf+>[,NRf+][,NRf+]
This command is used to set and query the slope of the frequency in normal
mode.
<NRf+>: frequency setting value.
Syntax:
FREQuency:SLOPe[:IMMediate] <NRf+>[,NRf+][,NRf+]
Query syntax:
FREQuency:SLOPe[:IMMediate]? <phase>
*RST:
1000Hz/s
Example
FREQuency:SLOPe 300.0