STATus Commands
Copyright © ITECH Electronic Co., Ltd. 122
Returns
<NR1>
STATus:OPERation:PTRansition
Sets the value of the PTR (Positive-Transition) registers. These registers serve as
a polarity filter between the Operation Condition and Operation Event registers.
When a bit in the PTR register is set to 1, then a 0-to-1 transition of the
corresponding bit in the Operation Condition register causes that bit in the
Operation Event register to be set. STATus:PRESet sets all bits in the PTR
registers and clears all bits in the NTR registers.
Syntax
STATus:OPERation:PTRansition <NR1>
Arguments
0-0x7fffffff
Default value
0
Example
STATus:OPERation:PTRansition 128
Query syntax
STATus:OPERation:PTRansition?
Returns
<NR1>
STATus:PRESet
This command is used to preset the value of the status register.
Syntax
STATus:PRESet
Example
STAT: PRES