(2) Load cable
As shown in the Fig. 2-2-1, please use hexagon nuts and hexagon head bolts when
you connect load cables. Also please use load cables adapted for necessary current
capacity and wire the load cables as short as possible. In addition, twist the load
cables if the cable length is long and/or if possible.
Fig. 2-2-3 How to fix load cables to the load terminals of this product.
Table 2-2-1 Suggested size for load cables
Never touch the Load terminals on front panel or rear panel while in operation.
Since front panel load terminals and rear panel terminals are connected
internally, voltage will appear when applied voltage to either on front or rear
panel terminals.
Never connect DUT at the same time to both on front panel terminals and rear
panel terminals.
Contrary will cause serious damage.
・ When reverse voltage is applied, the loading circuit of this unit will behave as
shorted and alarmed. But it will cause damage of internal circuit when applied
higher current than rating.
・ Never apply higher voltage than rating. When applied higher voltage, it will
alarm but may cause damage inside.