*RST — RESET Return 2016 to *RST defaults
Description
When the *RST command is sent, the Model 2016 performs the following operations:
1. Returns the Model 2016 to the *RST default conditions (see SCPI tables).
2. Cancels all pending commands.
3. Cancels response to any previously received *OPC and *OPC? commands.
*SAV — Save Save present setup in memory
Parameters
<NRf>=0
Description
Use the *SAVE command to save the current instrument setup configuration in memory for
later recall. Any control affected by *RST can be saved by the *SAV command. The *RCL com-
mand is used to restore the instrument to the saved setup configuration.
Only one setup configuration can be saved and recalled.
*SRE <NRf> — Service Request Enable Program service request enable register
*SRE? — Service Request Enable Query Read service request enable register
Parameters
<NRf>= 0 Clears enable register
1 Set MSB bit (Bit 0)
4 Set EAV bit (Bit 2)
8 Set QSB bit (Bit 3)
16 Set MAV bit (Bit 4)
32 Set ESB (Bit 5)
128 Set OSB (Bit 7)
255 Set all bits
Description
Use the *SRE command to program the Service Request Enable Register. Send this command
with the decimal equivalent of the binary value that determines the desired state (0 or 1) of each
bit in the register. This register is cleared on power-up.