Thermocouple commands
:TYPE <name>
[:SENSe[1]]:TEMPerature:TCouple:TYPE <name> Specify TC type
Parameters <name> = J Set operation for Type J thermocouples
K Set operation for Type K thermocouples
T Set operation for Type T thermocouples
Query :TYPE? Query thermocouple type
Description This command is used to configure the Model 2016 for the thermocouple
type that you are using to make temperature measurements.
These commands are used to configure the reference junction for
thermocouple temperature measurements.
:SIMulated <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:SIMulated <n>
Parameters <n> = 0 to 50 Specify temperature in °C
32 to 122 Specify temperature in °F
273 to 323 Specify temperature in K
DEFault 23°C, 73.4°F, 296K
MINimum 0°C, 32°F, 273K
MAXimum 50°C, 122°F, 323K
Query :SIMulated? Query simulated reference
:SIMulated? DEFault Query default *RST reference
:SIMulated? MINimum Query lowest allowable reference
:SIMulated? MAXimum Query largest allowable reference
Description This command is used to specify the simulated reference temperature. The
temperature value depends on which temperature scale is currently selected
(°C, °F or K). Typically, 0° or 23°C is used as the simulated reference
temperature.
5-48 SCPI Command Reference