:STATe <b>
[:SENSe[1]]:HOLD:STATe <b> Control (on/off) Hold
Parameters <b> = 0 or OFF Disable Hold
1 or ON Enable Hold
Query :STATe? Query state of Hold.
Description This command is used to enable or disable Hold. See Hold in Section 3 and
Trigger Model, Device Action in this section for details on Hold.
Speed commands
:NPLCycles <n>
[:SENSe[1]]:CURRent:AC:NPLCycles <n> Set NPLC for ACI
[:SENSe[1]]:CURRen[:DC]:NPLCycles <n> Set NPLC for DCI
[:SENSe[1]]:VOLTage:AC:NPLCycles <n> Set NPLC for ACV
[:SENSe[1]]:VOLTage[:DC]:NPLCycles <n> Set NPLC for DCV
[:SENSe[1]]:RESistance:NPLCycles <n> Set NPLC for 2
[:SENSe[1]]:FRESistance:NPLCycles <n> Set NPLC for 4
[:SENSe[1]]:TEMPerature:NPLCycles <n> Set NPLC for TEMP
Parameters <n> = 0.01 to 10 Power line cycles per integration
DEFault 1
MINimum 0.01
MAXimum 10
Query :NPLCycles? Query programmed NPLC value
:NPLCycles? DEFault Query *RST default value
:NPLCycles? MINimum Query minimum NPLC value
:NPLCycles? MAXimum Query maximum NPLC value
Description The integration period (measurement speed) for the basic measurement
functions (except Frequency and Period) is set using the :NPLCycle
command. NPLC (Number of Power Line Cycles) expresses the integration
period by basing it on the power line frequency. For example, for a PLC of
1, the integration period in seconds would be 1/60 (for 60Hz line power)
which is 16.67msec.
NOTE :SENS:VOLT:AC:NPLC and :SENS:CURR:AC:NPLC are valid only for a bandwidth
of 300Hz.
SCPI Command Reference 5-39