EasyManua.ls Logo

Keithley 2016 - Route Subsystem

Keithley 2016
269 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OUTPut:MODE <FIXED | LIST>
Parameters FIXED One FIXED output
LIST LIST of sweep values
Query :MODE? Query the output mode.
Description Sets the output mode to source one FIXED output or a LIST of sweep points.
LIST mode turns INIT:CONT OFF. Fixed mode sets SAMP:COUN 1 and
turns INIT:CONT ON. Pressing the LOCAL key sets the output mode to
FIXED and turns INIT:CONT ON. Saving a USER setup also sets the mode
to FIXED.
ROUTe subsystem
The commands in this subsystem are used to congure and control switching
and are summarized in Table 5-6.
:EXTernal <list>
:ROUTe:SCAN:EXTernal <list> Dene external scan list
Parameter <list> = (@ scanlist)
where scanlist is the specied list of external channels (1 to 800) to be
scanned.
Query :EXTernal? Query programmed scan list
Description The Model 2016 can operate with an external switch system, such as the
Keithley Model 7001 or 7002. The Model 2016 can measure up to 800
channels that are switched by the external switching system. This command
is used to dene the external scan list.
The scan list can contain 2 to 800 channels. The external scan is enabled by the
ROUTe:SCAN:LSELect EXTernal command.
:LSELect <name>
:ROUTe:SCAN:LSELect <name> Perform specied scan operation
Parameters <name> = EXTernal Enable scan for external scanner card
NONE Disable all scan operations
Query :LSELect?: Query scan operation
Description This command is used to select and perform the desired scan operation.
EXTernal is used to measure channels that are controlled by an external
switch system. When EXTernal is selected, the Model 2016 scans the
external scan list (see :SCAN:EXTernal). When NONE is used, the
Model 2016 disables all operations associated with a scan.
5-36 SCPI Command Reference

Table of Contents

Related product manuals