Measuring resistance
The Model 2016 can make 2-wire and 4-wire resistance measurements from 100µ to
120M.
Connections
Assuming factory default conditions, the basic procedure is as follows:
1. Connect test leads to the Model 2016 as follows:
A. For 2-wire, connect the test leads to INPUT HI and LO.
B. For 4-wire, connect the test leads to INPUT HI and LO, and SENSE 4 WIRE
HI and LO. Recommended Kelvin test probes include the Keithley Models 5805
and 5806. Either the front or rear inputs can be used; place the INPUTS button in
the appropriate position.
2. Select the measurement function by pressing 2 or 4.
3. Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE and keys to select a
measurement range consistent with the expected resistance.
4. Connect test leads to the resistance as shown in Figure 2-6.
CAUTION Do not apply more than 1000V peak between INPUT HI and LO or
instrument damage may occur.
5. Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until a normal reading is displayed. Use the lowest possible range for the best resolution.
6. Take a reading from the display.
2-22 Basic Measurements