:DIODe command
:RANGe[:UPPer] <NRf>
[:SENSe[1]]:DIODe:CURRent:RANGe[:UPPer] <NRf> Select current range for diode test
Parameters <NRf> = 0 to 1e-3 Specify diode test current
Query [UPPer]? Query selected range
Description There are three current ranges available for the diode test: 10µA range,
100µA range, and the 1mA range. Range is selected by using this command
to specify the expected current for the diode under test. The instrument will
then select the appropriate range.
:CONTinuity command
:THReshold <n>
[SENSe[1]]:CONTinuity:THReshold <NRf> Specify threshold resistance
Parameters <NRf> = 1 to 1000 Specify threshold in ohms
Query :THReshold? Query threshold resistance
This command is used to specify the threshold resistance for the continuity
test. Continuity occurs when the measurement is less than or equal to the
specified threshold level.
:DISTortion commands
:TYPE <name>
[:SENSe[1]]:DISTortion:TYPE <name> Select measurement type
Parameters <name> = THD THD type
THDN THD+n type
SINAD SINAD type
Query TYPE? Query measurement type
Description This command selects the type of distortion measurement. Selecting SINAD
sets the units to dB. Trying to change units to percent generates a
-221 Settings Conflict Error. Units remain dB when changing back to THD
or THD+n, but can then be changed to percent.
SCPI Command Reference 5-49