Setting limit values
Use the following steps to enter high and low limit values:
1. Press the SHIFT-LIMITS keys to view the present HI1 limit value:
HI1:+1.000000
^
This value represents the absolute value of that function.
2. Use the or keys to move to the number field. Use the , , , and keys to
enter the desired value. Move the cursor to the rightmost position (^) and use the and
keys to move the decimal point.
3. Press ENTER to view the present LO1 limit value:
LO1:-1.000000
^
This value represents the absolute value of that function.
4. Enter the desired value for this low limit.
5. Press ENTER to view the present HI2 limits value:
HI2: +2.000000^
This value represents the absolute value of that function.
6. Enter the desired value for this high limit.
7. Press ENTER to view the present LO2 limit value:
LO2: -2.000000^
This value represents the absolute value of that function.
8. Enter the desired value for the low limit. Pressing ENTER returns to the normal display.
Measurement Options 3-19