MEASure command
:MEASure[:<function>]?
<function> = CURRent:AC AC current
CURRent[:DC] DC current
VOLTage:AC AC voltage
VOLTage[:DC] DC voltage
RESistance 2-wire resistance
FRESistance 4-wire resistance
PERiod Period
FREQuency Frequency
TEMPerature Temperature
DIODe Diode testing
CONTinuity Continuity test
DISTortion Distortion
Description This command combines all of the other signal oriented measurement
commands to perform a one-shot measurement and acquire the reading.
When this command is sent, the following commands execute in the order
that they are presented.
:ABORt:CONFigure:<function>:READ?
When :ABORt is executed, if continuous initiation is disabled, the
instrument goes into the idle state. If continuous initiation is enabled, the
operation re-starts at the beginning of the Trigger Model.
When :CONFigure is executed, the instrument goes into a one-shot
measurement mode. See :CONFigure for more details.
When :READ? is executed, its operations will then be performed. In general,
another :ABORt is performed, then an :INITiate, and nally a FETCh? to
acquire the reading. See :READ? for more details.
5-6 SCPI Command Reference