:HARMonic:MAGNitude? <start#,end#>
[:SENSe[1]]:DISTortion:HARMonic:MAGNitude? <start#,end#> Query magnitude of measured
harmonics
Parameters <start#,end#> =2 to 64 Specify starting and ending harmonic
Description This command queries the magnitude of the measured harmonics of the last
date acquired without triggering a new reading. Both parameters are
required, even if querying a single harmonic (e.g., :MAGNitude? 2,2).
Non-integer values are truncated. The query does not return the magnitude of
any harmonics above the number set with the :HARMonic[:UPPer]
command.
:THD?
[:SENSe[1]]:DISTortion:THD? Query THD reading calculated
Description This command queries the THD reading calculated from the last acquired
data without triggering a new reading.
:THDN?
[:SENSe[1]]:DISTortion:THDN? Query THD+n reading
Description This command queries the THD+n reading from the last acquired data
without triggering a new value.
:RMS?
[:SENSe[1]]:DISTortion:RMS? Query ACV RMS reading of input wave
form
Description This command queries the ACV RMS reading of the input waveform
calculated from the last acquired data without triggering a new reading.
STATus subsystem
The STATus subsystem is used to control the status registers of the
Model 2016. The commands in this subsystem are summarized in Table 5-8.
SCPI Command Reference 5-53