Query :CONFigure? Query the selected function.
Description This command congures the instrument for subsequent measurements on
the specied function. Basically, this command places the instrument in a
one-shot measurement mode. You then use the :READ? command to
trigger a measurement and acquire a reading (see :READ?).
When this command is sent, the Model 2016 will be congured as follows:
The function specied by this command is selected.
All controls related to the selected function are defaulted to the *RST values.
Continuous initiation is disabled (:INITiate:CONTinuous OFF).
The control source of the Trigger Model is set to Immediate.
The count values of the Trigger Model are set to one.
The delay of the Trigger Model is set to zero.
The Model 2016 is placed in the idle state.
All math calculations are disabled.
Buffer operation is disabled. A storage operation currently in process will
be aborted.
Autozero is set to the *RST default value.
All operations associated with switching cards (scanning) are disabled.
This command is asserted when the :MEASure? command is sent.
Program PRINT #1, output 16; :conf:volt:dc”‘Perform :CONFigure
operations.
SCPI Command Reference 5-3