:AUTO <b>
[:SENSe[1]]:CURRent:AC:RANGe:AUTO <b> Control auto range for ACI
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO <b> Control auto range for DCI
[:SENSe[1]]:VOLTage:AC:RANGe:AUTO <b> Control auto range for ACV
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO <b> Control auto range for DCV
[:SENSe[1]]:RESistance:RANGe:AUTO <b> Control auto range for Ω2
[:SENSe[1]]:FRESistance:RANGe:AUTO <b> Control auto range for Ω4
[:SENSe[1]]:DISTortion:RANGe:AUTO <b> Control auto range for distortion
Parameters <b> = 1 or ON Enable auto range
0 or OFF Disable auto range
Query :AUTO? Query auto range (on or off)
Description These commands are used to control auto ranging. With auto ranging
enabled, the instrument automatically goes to the most sensitive range to
perform the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that
manually selects the measurement range (:RANGe <n>). When auto range
is enabled, the parameter value for :RANGe <n> changes to the selected
range value. Thus, when auto range is disabled, the instrument remains at the
selected range. When a valid :RANGe <n> command is sent, auto ranging
disables.
SCPI Command Reference 5-41