5. Set digital filter for ACV.
The 2016 has a digital averaging filter to stabilize readings. The repeating filter takes the
set number of readings, averages, then updates the display. The moving filter throws out
the oldest reading, takes a new reading, then updates the display. The moving filter
updates the display faster, but takes longer for readings to stabilize.
Front panel:
Press FILTER.
When the display shows “010 RDGS”, select the number of readings to average (1 to
100) and press ENTER.
When the display shows “TYPE:MOVING AV”, select MOVING AV or REPEAT
and press ENTER.
The display will have the FILT annunciator lit to indicate the filter is on. Press
FILTER again to disable the filter.
Remote:
:sens:volt:ac:aver:tcon mov|rep ;select moving or repeating filter for ACV
:sens:volt:ac:aver:coun xxx ;set number of averaged readings
:sens:volt:ac:aver:stat on|off ;turn filter on or off
Basic Measurements 2-45