19-62 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
file system
You can use the file system functions to access files saved on the USB flash drive. A compatible non-
Keithley application may be required to navigate in the file system or to view a list the files available.
These commands interact as an fs logical instrument on the TSP Platform to share a file system from
any node over the entire TSP-Link network.
To allow for future enhancements, the root folder of the USB memory stick has the absolute path
"/usb1/". Both the forward slash (/) and backslash (\) are supported as directory separators.
format.data
Attribute The data format for data printed using the printnumber and printbuffer functions.
Default format.ASCII
TSP-Link
accessibility
This attribute cannot be accessed from a remote TSP-Link node.
Usage fmt = format.data -- Reads data format.
format.data = fmt -- Writes data format.
Set fmt to one of the following values:
1 or format.ASCII ASCII format.
2 or format.SREAL Single precision IEEE-754 binary format.
2 or format.REAL32 Single precision IEEE-754 binary format.
3 or format.REAL Double precision IEEE-754 binary format.
3 or format.REAL64 Double precision IEEE-754 binary format.
Remarks This attribute selects the data format used to print data values with the printnumber and
printbuffer functions.
The precision of the ASCII format can be controlled with the format.asciiprecision
attribute. The byte order of SREAL, REAL, REAL32, and REAL64 can be selected with the
format.byteorder attribute.
REAL32 and SREAL select the same single precision format. REAL and REAL64 select the
same double precision format. They are alternative identifiers. Selecting which to use is a matter
of preference.
The IEEE-754 binary formats use 4 bytes each for single precision values and 8 bytes each for
double precision values.
When data is written with any of the binary formats, the response message will start with “#0”
and end with a new line. When data is written with the ASCII format, elements will be separated
with a comma and space.
Also see format.asciiprecision, format.byteorder, printbuffer, printnumber
Example Selects the ASCII data format:
format.data = format.ASCII
fs.chdir
Function Sets the current working directory.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage fs.chdir(path)
path The path of the new working directory. This path may be
either absolute or relative to the current working directory.
Remarks An error is logged to the error queue if the given path does not exist.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals