Series 2600A System SourceMeter® Instruments Reference Manual Section 3: Basic Operation
Recalling a saved setup
Setups can be recalled from internal nonvolatile memory or a USB flash drive at any time. To recall
a saved setup:
1. Press the MENU key to access the main menu.
2. Select SETUP, then press ENTER.
3. Select the RECALL menu item, then press ENTER.
4. Choose one of the following:
• INTERNAL
•USB1
5. (INTERNAL only) Do one of the following:
• Select the user number (1 through 5), then press ENTER
•Select FACTORY to restore factory defaults, then press ENTER.
6. (USB1 only) Select the appropriate file and then push ENTER.
To select power-on setup
1. Press the MENU key to access the main menu.
2. Select SETUP, and then press ENTER.
3. Select POWERON, and then press ENTER.
4. Do one of the following:
• Choose FACTORY to load the original defaults
•Select USER NUMBER (1 through 5) to load a user preference.
5. Press ENTER.
6. Press EXIT to return to the main menu.
Saving user setups from a command interface
Saving and recalling user setups
The setup.save and setup.recall functions are used to save and recall user setups:
Restoring the factory default setups
The reset functions return the Series 2600A to the original factory defaults:
setup.save(n)
setup.recall(n)
Where:
n = 1, 2, 3, 4 or 5
-- Save present setup to nonvolatile memory.
-- Recall saved user setup from nonvolatile memory.
reset() or *rst -- Restore all factory defaults.
smua.reset() -- Restore Channel A defaults.
smub.reset() -- Restore 2602A/2612A/2636A Channel B defaults.
setup.recall(0) -- Restore all factory defaults.