EasyManua.ls Logo

Keithley 2600a series - V-Source Operating Examples

Keithley 2600a series
696 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-14 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 4: Source-Measure Concepts Series 2600A System SourceMeter® Instruments Reference Manual
Figure 4-8
V-Source operating examples
Current Limit
Load Line
Voltage Source
Load Line
V-Source (V
S
)
I-Meter
(I
M
)
5mA
10mA
10V
2kW DUT Load Line (R)
I
M
= V
S
/ R
V
S
= I
M
 R
= 10V/2kW
= 5mA
A) Normal V-source operation
Current
Limit
Load Line
Operating
Point
Voltage
Source
I-Meter
(I
M
)
10mA
10V
= (10mA) (800W)
= 8V
B) V-Source in compliance
8V
800W DUT Load Line (R)
V-Source (V
S
)
Operating
Point

Table of Contents