EasyManua.ls Logo

Keithley 2600a series - Noise Shield

Keithley 2600a series
696 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-14 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 2: DUT Test Connections Series 2600A System SourceMeter® Instruments Reference Manual
Figure 2-19
Model 2636A High-impedance guarding (non-floating)
Noise shield
A noise shield (see Figure 2-20) is used to prevent unwanted signals from being induced into the
test circuit. Low-level signals may benefit from effective shielding. The metal noise shield
surrounds the test circuit and should be connected to SMU LO as shown in
Figure 2-20.
Figure 2-20
Models 2602A and 2612A noise shield
DUT
HILO
Noise Shield
CHANNEL A
LO GLO HI G G G HI
SS
KEITHLEY Series 2600A
Noise shield
connected to
In/Out LO

Table of Contents

Related product manuals